Ductless Filtered Fume Hood
Complies with NFX 15 211:2009, ASHRAE 110:1995, and Chinese national standard JG/T385:2012. It is ISO9001 certified.
- Product Introduction
Product Overview:
The BC-D Ductless Fume Hood Series is designed to protect laboratory personnel during experimental operations, preventing them from inhaling toxic, harmful, pathogenic, or unknown substances. It effectively purifies the air within the fume hood and the laboratory, providing a safe environment for laboratory personnel.

Product Specifications:
| Product |
|
|
|
|
Model |
BC-DS800 |
BC-DM1275 |
BC-DL1600 |
|
External size W*D*H mm |
800*620*1245 |
1275*620*1245 |
1600*790*1245 |
|
Internal size W*D*H mm |
781*574*934 |
1256*574*934 |
1581*744*934 |
|
Filter size mm |
370*395*50 |
||
|
Pre-filter size mm |
370*395*21 |
||
|
Air handling capacity |
230 m³/h |
690 m³/h |
|
|
Face velocity |
0.4-0.6 m/s |
||
|
Voltage |
110 V-240 V |
||
|
Frequency |
50-60 Hz (No need for UPS) |
||
|
Max. Current |
2A |
||
|
Noise level |
40-52 dBA |
||
|
Power |
42 W (Single Phase) |
126 W (Single Phase) |
|
|
Operation hole type |
Trapezoid |
||
|
Configuration |
|||
|
Filters |
4 sets |
6 sets |
|
|
Pre-filters |
2 sets |
3 sets |
|
|
Fan |
1 unit |
3 units |
|
|
Epoxy resin countertop |
1 pc |
||
|
Lighting |
1 set of LED lighting |
||
|
Display |
7-inch LCD touchscreen |
||
|
Control system |
1 set |
||
|
VOC alarm system |
2 units |
||
|
Temp. & humidity alarm system |
1 set |
||
|
Power cord |
1 pc |
||
|
Remote monitoring system |
1 set |
||
|
Optional |
Filter: Organic filter, Inorganic filter, Ammonia filter, Formaldehyde filter, HEPA Lower cabinet, Stainless steel countertop, Anemometer Alarm: Ammonia Alarm, Formaldehyde Alarm, HCL Alarm |
||
Product Features:
1.Main material:
≥1.2mm galvanized steel sheet
Epoxy resin electrostatic spray coating
Durable, chemical-resistant, lead-free coating
High gloss and minimal corrosion and moisture impact
2.Front and Side Panels:
Main material: ≥6 mm acrylic sheet
Excellent weather resistance and chemical resistance
Aging-resistant, colorless, transparent, and visually comfortable
3.Temperature and Humidity Monitoring System:
Displays real-time temperature and humidity
Alarm parameter settings to ensure safe operation
4.Fan Monitoring:
Fan system failure alarm
Adjustable fan speed to meet different chemical handling requirements
5.Filter Saturation Alarm System:
Equipped with dual filters and dual VOC sensors
One sensor monitors indoor air quality
The other sensor monitors filter saturation
Saturation alarm value setting, alerts when exceeded
If the concentration remains above the set value for an extended period, the filter needs to be replaced
6.PSC Fan from the USA:
24V voltage
Stable performance, ultra-quiet, and spark-free
7.High-Efficiency Filtering System:
Arranges particles according to their size, compliant with ASTM standards, Effective for acidic and organic gases, with strong adsorption capacity.
Adopts HEPA filter, achieving 99.995% efficiency for particles larger than 0.3μm.
8.LED Lighting:
Power equivalent to 25W fluorescent lamp
Generates no heat, safe, and does not affect lab temperature
Energy-saving, environmentally friendly, and long lifespan
9.Epoxy Resin Countertop:
Excellent chemical stability and corrosion resistance
High impact resistance, high-temperature performance
Durable, does not delaminate, expand, or crack
Software Control Section:

1. 7-Inch Large Screen Display:
IoT smart HMI interface
Resolution: 1024x600, perfect visual system
2. Chemical Management:
Supports a list of no less than 100 chemicals
List includes serial number, chemical name, quantity, unit, etc.
Add, delete, modify the list on the device
Data synchronized to the cloud server
View current chemical list via WeChat
3. Remote Management via WeChat:
View historical VOC data and temperature/humidity data in graphs
View historical alarm records
Remote control of fan, lighting, and alarm lights
4. API Interface Available, supports third-party platform integration
5. Software Control System is tested by the China Software Testing Center
Advantages:
No Duct Installation Required:
Easy installation, no exhaust emissions, environmentally friendly
Optional Top Filter Module System:
Meets various experimental needs
Advanced Modular Filtering Technology:
Fully adsorbs and filters harmful gases and particulate dust
Energy Efficient:
No air conditioning energy consumption, highly energy saving
Convenient Mobility:
Allows for proximity storage, facilitating easy access and improving work efficiency.

Working Principle:

- Inlet Air:
Air is drawn directly from the laboratory into the fume hood, forming a negative pressure which creates a natural barrier to isolate users from chemicals
- Comprehensive Filtration System:
The fan system draws toxic and hazardous gases upward, filters them, and returns clean air to the room, ensuring chemical gases inside the cabinet do not pollute the indoor environment
- Clean Laboratory Air:
Configurable filters for different operations
Effectively filters toxic gases and purifies lab air
Removes chemical odors, enhancing lab cleanliness
- Purified Surrounding Air:
Continuous filtration efficiency helps purify indoor air
Hot Tags: ductless filtered fume hood, China ductless filtered fume hood manufacturers, factory, Chemical resistant wash basin, Wall mounted single spout faucet, Quiet operation fans for laboratory ventilation, Chemical safety cabinet with ductless filter, Industrial purification fans for clean air, Wireless gas control valve